URGENT MEDICAL DEVICE CORRECTION

MiniMed FlexTM Insulin Pump Missing Manufacturing Test Data

MiniMed FlexTM insulin pump missing manufacturer test data — Download

Product ImpactModel/CFN NumberSerial Numbers
MiniMed FlexTM Insulin PumpMMT-8072 See Appendix I

August 2026

Dear Valued Customer,

We are writing to inform you of an important safety recall regarding your MiniMed Flex Insulin Pump. Refer to Appendix I for serial number information. Your safety and well-being are our highest priorities, and we want to make sure you have all the information you need to continue managing your diabetes safely.

Issue Description: 
MiniMed identified an issue in which certain MiniMed Flex Insulin Pumps (MMT-8062) were released to customers with incomplete manufacturing testing.

Health risk:
As a result of incomplete testing, your pump may have the potential for lack of protection from moisture exposure, slow charging, disconnection with sensor or MiniMed App, reduced ability to detect occlusion, no audio or low audio volume for alarms/alerts, or insulin under-delivery. If your pump experiences these issues, it could potentially lead to delayed therapy or risk of hypoglycemia, hyperglycemia, diabetic ketoacidosis (DKA) or death.

As of August 20, 2026, three (3) complaints have been reported of pumps failing to turn on at first use, with no adverse events reported to date. This recall is being reported to the U.S. Food and Drug Administration.

Required Actions:
1. Stop using the Flex pump in your possession and revert to your back-up plan for insulin therapy.

2. Please contact us immediately at (844) 885-0259 so we can arrange to ship you a new Flex pump overnight. Upon calling, please say “Flex” when prompted to connect with an agent.  

3. Please acknowledge that you’ve read and understood this notification and have followed the actions listed by completing and returning the confirmation form, scanning the QR code below with your phone or tablet, or by visiting https://info.medtronicdiabetes.com/FA1591 to acknowledge communication.

Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail, or by fax as follows: 

As always, we are here to support you and ensure we deliver the highest quality products possible. If you have further questions or need assistance, please call the MiniMed 24-Hour Technical Support line at 1-800-646-4633, option 1.

 

Sincerely, 

MiniMed

Appendix I - Subject Products

Product name: MiniMed FlexTM insulin pump

Model/CFN Number: MMT-8072

Serial Numbers

FB00010465

FB00011357

FB00011553

FB00014455

FB00014472

FB00014766

FB00014822

FB00015700

FB00015793

FB00015794

FB00015796

FB00015797

FB00015799

FB00015801

FB00015828

FB00015839

FB00015977

FB00016511

FB00018580

FB00019307